Meter Turbidity Infrared TN480
Downloads
Description
The TN480 uses an infrared light source and 90° scattering method, which is compliant with the ISO 7027 method. The instrument is designed for water quality professionals, offering long-term reliability in high-accuracy turbidity measurement with an intuitive operating system, simple calibration, and GLP data management.
- The infrared light source is especially ideal for testing solutions with color.
- TruRead measurement mode automatically takes multiple consecutive readings, and calculate their average, minimum and maximum value, as well as displaying each set of data in a brief test report. TruRead is a much better way to determine the true turbidity of your samples, rather than relying on just one reading. It's especially suitable for sample solutions with rapid settling and continuous measurement changes.
- The kit comes with AMCO high-molecular polymer turbidity standard solutions (0, 20, 100, 400, 800 NTU), which are EPA approved primary standards that are safe, non-toxic and disposable, and have a guaranteed one-year shelf life. It is NIST traceable to particle size/distribution and optical density, and its inherent stability means it does not settle from suspension. Store the standards at room temperature. They must not be frozen.
- Zero turbidity error reminder: the instrument has zero-point calibration and zero-point error reminder to ensure high accuracy for low-turbidity solutions.
- Large TFT color screen with simple operation guide, supporting English, Spanish, and Chinese.
- 200 sets of data storage with USB data output to PC (Windows-based), real-time clock display, calibration date check, automatic power off, and low battery reminder.
- Everything comes in a rugged carrying case, including AMCO calibration solutions (0, 20, 100, 400, 800 NTU), 6 pieces of sample vials, USB cable, silicone oil, a microfiber wipe, PC-Link software flash drive, and user manual.
Technical Data:
Measurement Method | 90 Degree Measurement, ISO 7027 Compliant |
Light Source | Infrared |
Measuring Range | 0 to 1,000 NTU (automatic range switch) 0.01 to 19.99 NTU 20 to 99.9 NTU 100 to 1,000 NTU |
Accuracy | ≤ ±2% of reading+ stray light |
Repeatability | ≤ ± 1% of reading or 0.02 NTU(FNU) (the greater of the two) |
Resolution | 0.01/0.1/1 NTU |
Stray Light | ≤0.02 NTU |
Calibration Standard | AMCO Polymer or Formazin Solution:0, 20, 100, 400 and 800NTU |
Detector | Silicon photovoltaic |
Measuring Mode | Normal measurement and TruRead measurement |
Data Storage | 200 groups |
Data Output | USB to PC |
Calibration Record | Calibration date and time |
Display | TFT color screen |
Sample Vial | Φ25×60 mm,high borosilicate glass with lid |
Sample Vial Volume | 18 ml |
Power Supply | 3-AAA alkaline batteries |
Working Condition | Temperature: 0 to 50°c Relative humidity:0 to 90% at 30c, 80% at 40c, 0 to 70% at 50c |
Instrument Sealing Grade | IP67 |
Dimensions and Weight | Unit: 90 x 203 x 80mm | 385g With Case: 310 x 295 x 110mm | 1.5kg |
AMCO Polymer Standard Calibration Solutions:
Approved by US EPA and ASTM, AMCO high molecular polymer turbidity standard solutions are the best alternatives to Formazin standards in terms of shelf life, ease of use and safety concerns.
Formazin Solution | AMCO Polymer Solutions | |
Toxicity | Highly toxic, PPE is necessary when handling | Non-toxic |
Operation | Requires diluting, complicated operation. | No diluting needed, use directly |
Shelf Life | <2 NTU: 1 hour; 2 – 20 NTU: 12 – 24 hours; 20 - 400NTU: 1 month | 1 year |
Storage Condition | Avoid sunlight at low temperature | Avoid sunlight at room temp. |
Convenience | Easy to settle, requires flipping and mixing | Very stable, can be used directly |
Traceability | Non-traceable | NIST traceable |
Supplied With:
- TN480 Meter
- Calibration Standards
- Sample Cuvettes x 6
- Microfibre Cloth
- Silicone Oil
- Software Flash Drive
- USB Cable
- Manual
- Carry Case
Details
-
Pack Size:EACH
-
Warranty (Months):12